58
5.6 Setup: P
r
essu
r
e Senso
r
- SETUP - PRESSURE SENSOR
UP DOWN
SELECT
SELECT
PRESSURE: RANGE
UP DOWN
SELECT
DOWN
PRESSURE: UNITS
UP DOWN
SELECT
DOWN
PRESSURE: 1
0.0
ZERO
SPAN
EXIT
PRESS FILTER: 4
UP DOWN
SELECT
PRESSURE SENSOR: Is used to configu
re settings related to the SciPres Pressure Sensors.
The following can be configured; Range, Units, Source (control and alarms), Filter (signal noise) and
Zero / Span. The LabTec™ uses only one sensor port, choice is made with the Source setting below.
RANGE: Set the overall pressure range, the Default is 60.0 psi (4.14 bar), and is the maximum.
Changing this setting requires re-calibration of the sensor signals.
UNITS: Select between psi, bar, and kpa. Default = psi.
SOURCE: Controls the source pressure sensor for monitoring and alarms. Select between P1, P2 or
P3. Default = P1
ZERO: Used to Zerothe sensors to establish the zero offset cause by the circuitry and any
inherent hydraulic pressure in your tubing beyond the pump head. Connect the sensor, and remove
any pressure from the system. Select the sensor, P1, P2 or P3 and pressZero. You are prompted to
confirm the action. “Span” should not be used on a routine basis, as it sets the max input value.
PRESSURE: SOURCE
UP DOWN
SELECT
DOWN
PRESSURE SOURCE: P1
UP DOWN SELECT
PRESSURE RANGE: 60.0
UP DOWN SELECT
PRESSURE UNITS:
psi
UP DOWN SELECT
PRESSURE: ZERO
UP DOWN
SELECT
DOWN
PRESSURE: ZERO
UP DOWN SELECT