57
- PRINT - PARIT
5.4
Setup: Ethernet
Ethernet is currently non-functional and is reserved for future development.
5.5
Setup: Printer
DOWN
DOWN
- PRINT - TYPE
TYPE:
SEIKO
- PRINT - BAUD RATE
BAUD RATE:
9600
UP
DOWN
SELECT
UP DOWN
SELECT
UP DOWN
SELECT
UP
DOWN
SELECT
DOWN
- PRINT - DTOP BITS
STOP BITS:
1
- PRINT - PRINT DELAY
PRINT DELAY:
0
UP DOWN
SELECT
UP
DOWN
SELECT
UP DOWN
SELECT
UP
DOWN
SELECT
PRINTER: Allows the setting of RS-232 communication parameters needed for connection to a
Parker serial printer or to a PC for data collection. These settings control the output
for both the
Printer and USB Ports.
PRINT TIME: Controls how fast the LabTec™ sends data points in Minutes: Seconds. Max is 30:00 min-
utes; Min is 00:05 seconds. Default = 00:30 seconds.
TYPE: Allows use of two styles of small serial printers, Seiko, a thermal unit, and Starr, a dot matrix.
For all other printers, and PC communication, Seiko setting is used and is the default.
STOP BITS: Select 1 or 2. Default = 1.
PARITY: Select Even, Odd or None. Default = None.
WORD LENGTH: Select 7 or 8 Bits. Default = 8 Bits.
BAUD RATE: Select 300, 600, 1200, 2400, 4800, 9600, or 38.4. Default = 9600
PRINT DELAY: Used to slow down the output for printers with small buffers. Select 0 – 5 seconds.
Default = 0 seconds.
- SETUP - PRINTER
UP DOWN
SELECT
SELECT
- PRINT - PRINT TIME
PRINT TIME:
00:10- PRINT - WORD LENGTH
WORD LENGTH:
8 BITS
UP DOWN
SELECT
UP
DOWN
SELECT
UP DOWN
SELECT
UP
DOWN
SELECT