EasyManua.ls Logo

PassMark PM123 - Ramp Up and Slew Rate Measurement; T2 Min Slew Rate Specification

Default Icon
29 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Inline PSU Tester User’s Guide Page 9
Figure 5
Power-on Time (T1): The power-on time is defined as the time from when PS_ON# is
pulled low to when the +12VDC, +5VDC, and +3.3VDC outputs are within the
regulation ranges. The power-on time shall be less than 500 ms (T1 < 500 ms).
Rise Time (T2): The output voltages shall rise from ≤10% of nominal to within the
regulation ranges within 0.1 ms to 20 ms (0.1 ms ≤ T2 ≤ 20 ms).
PWR_OK delay (T3): PWR_OK is a “power good” signal. It should be asserted high
by the power supply to indicate that the +12VDC, +5VDC, and +3.3VDC outputs are
above the under-voltage thresholds. The PWR_OK time shall be asserted high within
300ms to 500ms after outputs are above the under-voltage thresholds (300 ms < T3 <
500 ms).
Power-down warning (T6): During power-down, the PWR_OK signal should be
asserted low at least 1ms before the +12VDC, +5VDC, and +3.3VDC outputs reach
their under-voltage thresholds (T6 ≥ 1ms).
Ramp up
There must be a smooth and continuous ramp of each DC output voltage from 10% to
90% of its final set-point within the regulation band. The smooth turn-on requires
that, during the 10% to 90% portion of the rise time, the slope of the turn-on
waveform must be positive. The tester monitors the slope and report it as “Positive
Slope”. Also, for any 5ms segment of the 10% to 90% risetime waveform, a straight
line drawn between the end points of the waveform segment must have a slope ≥
[Vout, nominal / 20] V/ms. The tester measures the minimum slope across all 5ms
segments during power-up and displays it as “T2 Min Slew Rate”.
Output
T2 Min Slew Rate (V/ms)
+12V1
0.6
+5V
0.25
+3.3V
0.165
+12V2 (12V CPU)
0.6
Table 8