NMS044 Reference Manual Shipping and Storage Requirements A-5
A.2 Shipping and Storage Requirements
The box and packing foam your system arrived with is designed to be reused. We
recommend that you retain the packaging for future shipments.
A.2.1 Shipping the System
If you need to ship your system, be aware that the LiFePo Battery (BAT019) is considered
Class 9 Hazardous Material. The shipping vendor and you are required by law to follow
specific protocol when shipping. One requirement is that a company and/or individual
must be 49 CFR and IATA certified to ship a lithium battery with over 100 Watt-hour
capacity. Recertification is required every 2 years
Prior to shipping the system, always disconnect the battery from the power block
terminal. The battery can be shipped in the case when you affix appropriate shipping
labels on the outside of the box.
CAUTION Do not transport with lid open. Only those with appropriate certifica-
tion may ship batteries.
LEARN MORE Licensing can be obtained through a training course, such as the
Lion Technology online training course - code #HMT 254 “Shipping Lithium
Batteries”.
Table A.5 (Cont) Genasun Solar Charge Controller LED Indication Patterns
...LED Status Indications
Battery Charged The battery is in the
absorption or float charging stage.
SOLID GREEN LED
...LED Error Indications
Panel Voltage Too High Only 12V nominal
solar panels may be used with this
controller.
SETS OF 6 RED BLINKS
Internal Error Contact your dealer
2 LONG BLINKS, THEN ANY NUMBER OF
SHORT BLINKS