EasyManua.ls Logo

PCB Piezotronics Larson Davis NMS044 - Page 55

PCB Piezotronics Larson Davis NMS044
59 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
NMS044 Reference Manual Configuring the Gateway for Larson Davis Instruments A-17
Step 10. Choose the Serial menu, open the Port Configuration list, select Disable from
the Serial Port drop-down menu, and click Apply.
Figure A-19 Serial Port Settings
Step 11.
Select the LAN menu, USB sidetab; verify that the settings are as shown in
Figure A-20, and click Apply.
Figure A-20 USB Port Settings
Step 12.
Select the I/O menu, Configuration sidetab; verify that the settings are as
shown in Figure A-21, and click Apply.
Figure A-21 Settings on the I/O Tab