Page 11
Technical Specification
Genius 1022
Environment
Minimum Operating Ambient Temperature 5°C (41°F)
Maximum Operating Ambient Temperature 30°C (86°F)
Maximum Altitude 3000 m
Maximum Relative Humidity 80% Non-Condensing
Minimum Storage Temperature* -20°C (-4°F)
Maximum Storage Temperature* 60°C (140°F)
*NOTE – When taken out of storage the Generator should be allowed to acclimatize at room temperature
for a minimum of 3 hours before operation.
Generator Outlets
Maximum Gas Output Pressure 7.9 bar (116 psi)
Maximum Outlet Flow Rate 32 L/min (1.13 cfm)
Particles <0.01µm
Phthalates NONE
Suspended Liquids NONE
Gas Outlets 1 x ¼” BSPP
Drain outlet 1 x ¼” BSPP
Pressure gauges 1
Electrical Requirements
Voltage 220-240V ± 10%
Frequency 50/60 Hz
Current 7 Amps
Input Connection C20 Plug
Power Cord (Supplied) C19 socket to local connection (13A minimum)
Pollution Degree 2
Insulation Category II
General
Dimensions cm (inches) WxDxH 60 x 75 x 71.2 (23.6 x 29.5 x 28)
Generator Weight Kg (lbs) 95 (209)
Shipping Weight Kg (lbs) 118 (260)