EasyManua.ls Logo

Peak Scientific NM32LA-A - Genius NM32 LA;NM32 LA-A Technical Specifications

Default Icon
33 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Page 13
Technical Specification
Genius NM32LA/NM32LA-A
Environment
Minimum Operating Ambient Temperature 5°C (41°F)
Maximum Operating Ambient Temperature 30°C (86°F)
Maximum Altitude 3000 m
Maximum Relative Humidity 80% Non-Condensing
Minimum Storage Temperature* -20°C (-4°F)
Maximum Storage Temperature* 60°C (140°F)
*NOTE – When taken out of storage the Generator should be allowed to acclimatize at room temperature
for a minimum of 3 hours before operation.
Please ensure generator is situated in a well ventilated environment
Generator Outlets
Maximum Gas Output Pressure 6.9 bar (100 psi)
Maximum Outlet Flow Rate 32 L/min (1.13 cfm)
Particles <0.01µm
Phthalates NONE
Suspended Liquids NONE
Gas Outlets 1 x ¼” BSPP
Drain outlet 1 x ¼” BSPP
Pressure gauges 1
Electrical Requirements
Voltage 220-240v +/- 10%
Frequency 50/60 Hz
Current 7 Amps
Input Connection C20 Plug
Power Cord (Supplied) C19 socket to local connection (13A minimum)
Pollution Degree 2
Installation / over voltage Category II
General
Dimensions cm (inches) WxDxH 60 x 75 x 71.2 (23.6 x 29.5 x 28)
Generator Weight Kg (lbs) 102.5 (226)
Shipping Weight Kg (lbs) 128 (282)