Maintenance
103
3. Replace the spent fuse with a new one of the same type and rating:
Voltage Standard
100 V – 120 V 6.3 A T
210 V – 240 V 3.15 A T
NOTE: The module has two fuses. Although only one fuse may have blown both fuses should
be replaced at the same time.
4. Replace the fuse holder.
Align the lug at the bottom of the fuse holder with the slot in the socket. A
click is heard as each lug snaps into place.
NOTE: If you use the correct fuses but the instrument still does not work correctly, or the
fuses blow repeatedly, contact your PerkinElmer office or representative.