09931286E
19
Measurement
Thermal Denaturation Mode
1. Install the 6-Position Peltier Controlled Cell Changer referring to the chapter Installation.
2. Close the sample compartment cover and turn on the instrument.
NOTE:
Before System Self Test is performed, the 6-Position Peltier Controlled Cell Changer should be installed.
Otherwise, the instrument can be damaged electrically and do not operate properly.
NOTE:
Start the System Self Test after warming up the system for at least 20 minutes.
3. Double-click on the UV Express folder and select Thermal Denaturation mode.
4. Select OK after finishing the System Self Test.
5. Select New to open a new window and select OK.
NOTE:
When the 6-Position Peltier Controlled Cell Changer is connected for the first time, you need to correct
the cell position. Calibrate the beam position of the 6-Position Peltier Controlled Cell Changer referring to
the chapter Calibration of Multi-Cell Position (p30).
6. Open the method in main menu or mode icon on the left side of the main window. Set parameters
for Experiment.