b. Using Cells
Show the cell positions which are selected for measurement. Remove the cells by and
switch positions by and keys.
c. Multi Zero, Multi Baseline
If checked, zero (baseline) will be measured all the selected positions. If not, zero (baseline) will
be only measured at the 1
st
cell position among the using cells. Multi Baseline is only available at
the Scan and Scanning Kinetics mode.
NOTE:
To prevent inadequate baseline (zero) measurement because of the difference in transmittance among
cuvettes or any other reasons, Multi Baseline (Multi Zero) measurement is recommended.
d. Temperature (°C): Enter the returning temperature for the experiment.
e. Monitor
Select temperature that used to monitor during measurement. Options include: Block, Probe1 or
Probe2.
f. Error Range (°C/min)
It shows the temperature tolerance between the sampling and monitored temperature. The
measurement will start when the temperature tolerance reaches within the set Error Range.
Example: Set the starting temperature at 20°C and error range with 1°C then, the measurement
will start when the temperature is reached between 19°C and 21°C.
▶ Temperature Display
g. Select which temperature will be displayed on the panel:Block, Probe 1 or Probe 2. The only
selected temprature is displayed in real time.
h. Offset:
This function is used to adjust temperature. It is only used for manfacturing process, so
do not modify the values.
10. After completing the parameter setup, select Apply and OK. Then the LED of ON-LINE is turned on,
it will start heating up or cooling down to the set temperature in the Setup Multi Cell Peltier tab.