6-Position Peltier Controlled Cell Changer does not move
1. Check if the 6-Position Peltier Controlled Cell Changerconnector is connected to the Lambda 365.
2. Turn on and off again the Lambda 365 with installed 6-Position Peltier Controlled Cell Changer.
Intensity is too low
1. Recalibrate if the light beam does not reach the center of the cell holder’s hole.
2. If the intensity value is still low after the recalibration, the lamps of the light source may be
deteriorated and need replacing. Contact Perkin Elmer or the nearest representative.
When 6-Position Peltier Controlled Cell Changer does not reset
1. If reset is failed, the photo interrupt switch inside the 6-Position Peltier Controlled Cell Changerneeds
replacing. Contact Perkin Elmer or the nearest representative.
When 6-Position Peltier Controlled Cell Changer is not aligned precisely
1. Recalibrate the 6-Position Peliter Controlled Cell Changer if the light beam does not reach the center
of the cell holder’s hole.