Why after "Vent" control commands vacuum pumps continue running on Pfeiffer Vacuum GSD 350 OMNISTAR?
- BbsullivanSep 12, 2025
If, after the "Vent" control commands, the vacuum pumps continue running and the system is not vented, it could be because the delay time for venting has not yet completed. Wait for the delay time of less than 15 minutes to complete. The time is shown at the top right of the display.