Operating Instructions
Blood Bank Refrigerator

Questions and Answers

  • T
    Tracey MooreAug 11, 2025
    Why is the chamber of my Phcbi Refrigerator not getting cold enough?
    • K
      Kristine ScottAug 11, 2025
      If your Phcbi Refrigerator chamber isn't getting cold enough, consider the following possible causes: * The outer door is opened frequently or the inner door is open. * The unit is in direct sunlight, there is a nearby heat source, or the ambient temperature is too high (the allowable ambient temperature is between 5 °C and 35 °C). * Ventilation around the unit is blocked. * Too many items are stored inside, or the air exhaust vent is blocked by stored items. * The access port is not covered and should be covered with insulation and rubber caps when not in use. * The door gasket is damaged. If this is the case, contact our sales representative or agent for replacement. * A foreign substance is located between the door gaskets.
  • M
    Matthew BrownAug 21, 2025
    Why is my Phcbi Refrigerator making excessive noise?
    • A
      Angela NewmanAug 22, 2025
      Excessive noise during the operation of your Phcbi Refrigerator can be attributed to: * The floor not being sturdy. * The installation site not being level. * The refrigerator being tilted. * The cabinet touching the surrounding wall.
  • J
    Jose WilliamsSep 1, 2025
    Why is the displayed temperature not 4 °C on my Phcbi Refrigerator?
    • G
      Gary GregorySep 1, 2025
      A displayed temperature between 3 °C and 5 °C (depending on the ambient temperature) is not a malfunction.

Phcbi MBR-1405GR Specifications

General IconGeneral
BrandPhcbi
ModelMBR-1405GR
CategoryRefrigerator
Temperature Range2°C to 14°C
Temperature Uniformity±3°C
Alarm TypesHigh temperature, Low temperature, Door open, Power failure