EasyManua.ls Logo

Plusmed pM-100 fasTTest - The Control Solution Test; When to Perform a Control Solution Test

Default Icon
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
74
The Control Solution Test
When to Perform a Control Solution Test
The Control Solution is used to check and ensure that the plusmed fasTTest Blood
Glucose Monitoring System and plusmed fasTTest Test Strip are working together
properly and that you are performing the test correctly.
1. When you are using your Blood Glucose Monitoring System for the rst time.
2. When you are using a new batch of Test Strips.
3. Anytime you question the performance of the system, or on a regular basis to
ensure accuracy, e.g. once a week.
4. When you adjust your diabetic medication plans.
5. When your blood glucose test result is lower or higher than your normal level.
Note:
There are three Control Solutions at different desired range of aqueous glucose
(Low, Normal or High). For more information about the Control Solution, please read
your Control Solution package insert. Two sets of control solution should be used, a
high level and a low level control solution of your choice. For order information, please
contact your dealer.