35
Version 2.1
Update: April 2016
PONSEL Mesure Aqualabo Contrôle
35 Rue Michel Marion 56850 CAUDAN, FRANCE
Tel.: +33(0)2 97 89 25 30 Fax: +33(0)2 97 76 55 72
For a Voltage input: can be set to a value between the minimum value set
up in the "Signal min" line and 10 V.
I2: Average
Used to set the number of measurements, from 1 (instantaneous value) to
50, from which the moving average is calculated.
I2: Refresh rate
This line lets the operator select the interval between each measurement.
The possible values are:1 to 60 seconds.
I2: Error reporting
Inactive (no error reporting)
Out-of-range alert: if the measurement falls outside of the range
programmed by the Signal min and Signal max lines, information is sent to
the relay output if the programming is set to "Alarm mode" (refer to section
6.2.6).
I2: Source signal 1
units
----------- : units not selected
mH
2
O: water height measurement sensor
mbar: pressure sensor
bar : pressure sensor
I2: Source signal 1 min
This line lets the operator set the minimum value to between -9999.99 and
9999.9
I2: Source signal 1
max
This line lets the operator set the maximum value to between the minimum
value set on the "Source signal 1 min" line and 9999.9.
I2: Source signal 1
display
This line lets the operator arrange the information displayed on the home
screen (refer to Table 2 for parameter display details)
Not displayed
Sens 1. Prim. Meas: the measurement shall be displayed in the space
allocated to the primary parameter for sensor 1 (item 3 on the home screen)
Sens 1. Sec. Meas.: item 9 on the home screen
Sens 2. Prim. Meas: item 6 on the home screen
Sens 2. Sec. Meas.: item 13 on the home screen.
I2: Source signal 1
desig.
------------ : no designation
Water height
Defined by User: in this case, an additional line will appear as soon as this
option is confirmed.
I2: User desig.
A free text field for entering the user's designation (maximum of
13 characters).
I2: Source signal 1
stability delta
Can be set from 0 to 50 % (the default is 1 %) and is used to define when
the arrow on the home screen will indicate that the measurement is stable if
the parameter measured has been set as the primary parameter for sensor
1 or 2 (refer to items 2 and 5 in Table 2)
I2: Source signal 2
Inactive
Active: source signal 2 is calculated from source signal 1 by a conversion
program written in a file stored on the USB stick plugged into the USB port.
The name of the file must be of the form "CONFANA2.csv" and must
contain the following information:
Line 1 of the csv file: a description (not more than 16 characters) of the
calculated source signal.
Line 2: a blank line containing no characters