2.2.3 EHAN sensor: Redox and temperature.
The PONSEL sensor incorporates a reference electrode, used for ORP measurements, such as Ag /
AgCl in saturated KCl electrolyte plasticized "PLASTOGEL" ®.
The electrolyte "PLASTOGEL" ® communicates directly with the external environment without
interposition of capillary or porous. So there is no risk of clogging or defusing the reference.
The measuring electrode is in platinum (3,5mm2) presented in sealed ring on a glass rod and is for in-
situ measurements continuously
Temperature: measures via CTN.
Combination Electrode (ORP/reference) platinium
ring, Reference Ag/AgCI. Gelled electrolyte (KCI)
Modbus RS-485 standard and SDI-12 in option
Standby : 25 µA
Average RS485 (1 measure/second) : 20 mA
Pulse current : 500 mA
Heating time : 100 mS
Dimensions of sensor mounted
Mounted sensor length: gland (262 mm) not
included ;
Length with gland : 324 mm.
Material in contact with the environment
PVC, POM-C,platinum, Polyurethane
9-wire shielded conductor, uncoated-strand
polyurethane sleeve