www.posey.com 1.800.447.6739
KeepSafe Essential
14
To Change/Select Mode:
1. Ensure alarm is on and magnet is on
face plate.
2. Remove the battery compartment door
by sliding down to gain access to the
programming buttons.
3. Slide the MODE Select Switch to the
desired alarm mode (Fig. 4).
The last option selected is the mode
utilized when the magnet is removed
from the face plate.
4. Replace the battery door and slide shut, locking into place.
NOTE: Cat. 8373NP does NOT have an ON/OFF switch. To set Alarm
Mode for Cat. 8373NP, follow all steps above except step 1.
THE PRE-RECORDED VOICE MESSAGE IS:
“Please don’t get up. Sit back down and use the call button to call for help.”
Fig. 4
Tone Volume
Voice
&
Tone
VoiceTone
Qty 4AAA Alkaline
ANSI-24A IEC-LR03
Setting Alarm Mode
The KeepSafe Essential has three (3) alarm modes. These allow you to
select an alarm signal best suited to patient/resident and facility needs.
SETTING MODE
Mode Features
VOICE &
TONE
Pre-recorded voice message plays once, followed by
the selected alarm tone. The tone continues to play
until alarm is placed on suspend or patient/resident is
repositioned with the magnet on the faceplate.
TONE
Selected alarm tone plays until alarm is placed on
suspend or patient/resident is repositioned with the
magnet on the faceplate.
VOICE
ONLY
Pre-recorded voice message plays until alarm is placed
on suspend or patient/resident is repositioned with the
magnet on the faceplate.