36
BATTERY MAINTENANCE
Battery Maintenance
Battery Compartment
If there is ANY evidence of battery leakage, remove
the alarm from use and notify the appropriate facility authority.
The alarm should be disposed of according to your facility disposal
requirements. DO NOT use the alarm and DO NOT attempt to clean
it if there are any signs of battery leakage such as corrosion, rust
or white powder residue.
DO NOT allow batteries to deplete while in the
alarm. Change batteries immediately when hearing the low
battery “chirp.” Depleted batteries may leak and corrode,
causing damage to the electronics and reliability. When storing
the alarm with power “on”, check the alarm every week to
make sure the batteries are still operable and the alarm is still
on. If the alarm low battery alert is chirping, or the alarm does
not power up, the batteries are depleted and must be removed.
DO NOT leave depleted (“dead”) batteries in the alarm to avoid
corrosion.
Remove batteries when storing the alarm for an extended period
to prevent depleting the batteries and potential corrosion.