Monitoring in Cord and Magnet Mode (Cont.)
CHAIR USE
For use with an adjustable chair, lengthen or shorten cord to account
for any change in distance when chair is adjusted.
Make sure cord is not impeded by chair cushions and is not entangled
in chair parts.
TESTING MAGNET CORD
ALWAYS check magnet when con
-
necting it to a Posey alarm. You can
check a magnet by removing it from the
face plate of the alarm (Fig. 20). When
the magnet is disengaged, the alarm
should sound. DO NOT use the alarm or
magnet if it does not activate each time
magnet is removed from face plate.
ALWAYS check that magnet cord is not
frayed or worn and could break. DO NOT
use if cord is damaged.
H
O
LD
/
SUSPEN
D
FIG. 20