EasyManuals Logo

POSITECTOR SST User Manual

POSITECTOR SST
21 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #7 background imageLoading...
Page #7 background image
Page 5
Setup
Set configuration options for the Gage.
Patch Volume
Set the volume of water being used for the test (3.0 ml default). Use the (-) and (+)
buttons to decrease/increase the displayed value. The patch volume is used to calculate
surface density.
Set Clock
All measurements are date and time stamped (24 hour format) when stored into
memory. It is therefore important to keep both the date and time current using this menu
option. Use the Up and Down buttons to scroll, and the (-) and (+) buttons to adjust
value. The Gage’s date and time can also be viewed in Gage Info and on top of the
main menu.
NTP
Enables automatic time synchronization with a remote server over WiFi using the NTP
(Network Time Protocol).
Enter a NTP Server address and TZ (Time Zone) details. The PosiTector uses POSIX
format to set the time zone.
Eastern Time in the USA is input as the following:
EST5EDT4,M3.2.0,M11.1.0
From left to right:
- EST5 is the time zone name and offset from UTC
- EDT4 is the daylight savings time zone name and offset from UTC
- M3.2.0 is the start of EDT. ‘M’ specifies that the date will follow a ‘Month.Week.Day’
format. 3.2.0 is the third month, second week, first day (Sunday)
- M11.1.0 is the end of EDT. ‘M’ specifies that the date will follow a ‘Month.Week.Day’
format. 11.1.0 is the eleventh month, first week, first day (Sunday)
Detailed NTP instructions available at: www.defelsko.com/NTP