■ 9 Error messages and correction of faults
50
9 Error messages and correction of
faults
9.1 Error messages
The following table shows error messages and their possible causes.
If you cannot clear the error on the basis of the table, please contact
a Precisa Service Engineer.
Error
Possible cause
Err 1 • Change of application:
Invalid application selected
Err 3 • Count, Percent:
Weight too small
Err 4 • Anti-theft encoding:
Code not correct
Err 5 • Anti-theft encoding:
Code not correct
Err 8 • Switching on:
Zero point out of range
Err 9 • Calibration:
Calibration bad (repeat calibration)
Err 31 • Calibration:
No weight defined
Err 47 • Statistics:
Too much values
Err 48 • Statistics, Net-Total, Summation:
Different unit as for the first value
Err 49 • Statistics, Net-Total, Summation:
Too few values (min. 2 for Statistics)
Err 53 • Statistics:
Value out of range (+/-50%)
• Animal weighings:
Balance has been in „OL“ or „UL“