14 Further information ■
61
14.2.1 External calibration by means of ICM
Depending on the type of balance, calibration weights in steps of
10 g, 50 g, 100 g and 500 g can be used, where the calibration
weight must correspond to the precision of the balance.
For an external calibration by means of ICM, „ModE EXT“ must be se-
lected in the configuration menu (see chapter 5.5 “Calibration func-
tions”).
• Switch to „bALAncE” with the «MODE»-key.
CAL • Press «T» until „CAL” appears.
- - 0000 9
• The balance carries out a Zero measurement
(- - 0000 g is shown flashing).
- - 200 9
• After the zero measurement the display
flashes with the recommended calibration
weight.
• Place the calibration weight on the pan.
- - 200 9
• The display continues to flash.
• Calibration is complete when the display stops
flashing.
+ 200 9