9.4 Repetitive Measurement
CAUTION
Risk of burns!
If a measurement has already been carried out by means of the Pocket-
Monitor, the absorber can still be very hot (depending on the absorber
version for a duration of up to 30 minutes).
Only touch the display- or operating unit or the swivel joint of the device.
A new measurement only makes sense if the absorber has cooled down and its tem-
perature equals the ambient temperature.
In order to cool down the absorber, it has
to be submerged in cold water.
Do not submerge the display- or operating
unit or the swivel joint!
Press
TEMP
.
Observe the temperature display. As soon
as absorber temperature approx. equals
the ambient temperature (or exceeds it by
a maximum of 5°C), the absorber can be
taken out of the water.
Notice: If the display reaches the negative
range quickly, it is still too hot due to the
prior measurement and has to cool down
for a longer period of time.
Carefully dry the absorber by means of
compressed air. This cools it down to the
ambient temperature and makes it ready
for the next measurement cycle.
Proceed as described in chapter „9.3
Single Measurement“ on page 18.