12 Current Loop Interface (OPTION)
This version of the separate display- and operating unit has an additional 4-20mA-current
output. The power is given out by means of an impressed current. A power of 0 Watt
equals a current of approx. 4mA. The current of a maximum of 20mA equals 16bits.
Current Loop Interface
Fig. 12.1: Display- and operating unit with current loop interface
Socket, 5-pole
View: Plug Side
Pin Function
1
2
5
4
3
1 GND
2 Supply voltage +24V, max. 30mA
3 Current output 4 ... 20mA (maximum load of 500ohm)
4 Start impulse for 20s measurement (+24V regarding GND)
5 Start impulse for 10s measurement (+24V regarding GND)
When starting the measurement, i.e. when the timer button is pressed, the current
temperature is sent to the current loop interface. The temperature can be measured by
means of the following formula:
Type Calculation Formula
PMT 01p; PMT 70icu T = (I − 4mA) • 13.652 °C/mA
PMT 120icu T = (I − 4mA) • 27.307 °C/mA
PMT 002p; PMT 05p; PMT 30p T = (I − 4mA) • 6.827 °C/mA
T = Temperature in Degree Celsius
I = Impressed Current in mA