EasyManua.ls Logo

Primes PocketMonitor - Power Measurement Formula; Measurement Initiation and Operation

Default Icon
35 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
35,0(6
PocketMonitor PMT
24
Revision 01/2017 EN
After the irradiation time (= laser beam on the absorber) has run out, the output is set
back to 4mA. The measurement of this current can be used for the correction of osets.
In this case, the measured value is used in this formula with a set current of 4mA.
After the thermalization time has run out (depending on the absorber model 10...60sec-
onds) the power is given out.
Type Calculation Formula
PMT 002p P = (I − 4mA) • 3.4134 W/mA
PMT 01p P = (I − 4mA) • 6.8268 W/mA
PMT 05p P = (I − 4mA) • 34.134 W/mA
PMT 30p P = (I − 4mA) • 341.34 W/mA
PMT 70icu P = (I − 4mA) • 682.68 W/mA
PMT 120icu P = (I − 4mA) • 1365.36 W/mA
P = Power in watt
I = Impressed current in mA
The 20s- or 10s measurements can be started via the 24V inputs. The PocketMonitor
is supplied with current by means of the 24volts for the current loop.
With regard to the external supply of the PocketMonitor with 24V, the device does not
turn itself off automatically but is constantly in operation. The measurement can be
started any time by means of a start impulse for pins 4 or 5.
The temperature can be measured by means of the start impulse.