After the irradiation time (= laser beam on the absorber) has run out, the output is set
back to 4mA. The measurement of this current can be used for the correction of osets.
In this case, the measured value is used in this formula with a set current of 4mA.
After the thermalization time has run out (depending on the absorber model 10...60sec-
onds) the power is given out.
Type Calculation Formula
PMT 002p P = (I − 4mA) • 3.4134 W/mA
PMT 01p P = (I − 4mA) • 6.8268 W/mA
PMT 05p P = (I − 4mA) • 34.134 W/mA
PMT 30p P = (I − 4mA) • 341.34 W/mA
PMT 70icu P = (I − 4mA) • 682.68 W/mA
PMT 120icu P = (I − 4mA) • 1365.36 W/mA
P = Power in watt
I = Impressed current in mA
The 20s- or 10s measurements can be started via the 24V inputs. The PocketMonitor
is supplied with current by means of the 24volts for the current loop.
With regard to the external supply of the PocketMonitor with 24V, the device does not
turn itself off automatically but is constantly in operation. The measurement can be
started any time by means of a start impulse for pins 4 or 5.
The temperature can be measured by means of the start impulse.