Approximately 100 measurements can be recorded for
each test surface. The interval between individual mea-
surements must be at least 10 mm.
During checks the test block must be
placed on a solid, flat surface.
13.1 Viewing the data in the memory
Simple operation of the electronic system was extended
by an important function; «View memory» by pressing a
key. All measurements in the memory can be shown on
the LCD display and scrolled through for viewing.
The data in the memory can only be viewed when a
measurement series has been carried out. To finish a
measurement series, press the «END» key.
Press the i key in the last measurement series to
access memory mode.
The measurement profile of the viewed measurement
series and the corresponding statistics is shown on the
display ( Fig. 73 ). The arrow «i» on the display shows
the end of the memory. The designation «MEM» shows
that memory viewing is active. No measurement is pos-
sible in this mode «memory = MEM».
- Recommended charging time approx. 15 hours,
up to 24 hours max.
- When the printer is not in use, always switch the
switch to the OFF position.
13 Other important information
PAROTESTER2 has an automatic deactivation feature.
The electronic system switches off automatically after
60 minutes.
PAROTESTER2 always monitors the status of the bat-
teries and signals «Change batteries» on the display.
The electronic system of the PAROTESTER2 constant-
ly monitors error-free function of the impact device.
Deviations are signaled by messages on the display
(also refer to section 10.0 «Error messages»).
If the ON/OFF button is actuated during a series of
measurements, the data of the started measurement
series are lost.
Striped bars on the display or the printout mean that
the measured values are above or below the selected
display range.
A calibrated test block type U is available as an acces-
sory for periodical checks of the PAROTESTER2.
Important information 31