Moisture: 3.6% 01
Dual: Y/N
Dual Mode
The MCT May be switched Single
to Dual Mode. The triple mode
requires a different EPROM to be
fitted to the sensor.
Moisture: 3.6% 01
Language: Y/N
Language Selection
The MCT display system may be
set to use a different language to
English. The EPROM fitted to the
MCT will allow the language to be
switched betweeen English and
one of the following languages:
German, French, Spanish or
Portuguese
Moisture: 3.6% 01
Dry: Y/N
Dry Basis
The MCT may be setup to present
moisture readings as a % of the dry
material. Selection of Y will
activate this 'Dry basis' readout.
Refer to Appendix III for greater
details of this 'Dry Basis' readout.
Moisture: 3.6% 01
Max Cal: 10
Max Cal
The number of Calibration
Channels may be set to any
number between 1 and 10. Move
the cursor over the number set
desired value. Press ENTER to
save the value.
51