3302F Mainframe Operation Manual 49
OCP: STEP
Command Syntax:
[PRESet] OCP:STEP {SP}{NR2}{NL}
[PRESet] OCP:STEP {NL}
Purpose: Set and read the increasing value of OCP Test
Description: This command is used for setting the increasing value (I-STEP) of OCP
Test.
OCP: STOP
Command Syntax:
[PRESet] OCP:STOP {SP}{NR2}{NL}
[PRESet] OCP: STOP? {NL}
Purpose: Set and read the maximum value of OCP Test
Description: This command is used for setting the maximum value (I-STOP) of OCP
Test.
VTH
Command Syntax:
[PRESet] VTH {SP}{NR2}{NL}
[PRESet] VTH ? {NL}
Purpose: Set and read the value of the Threshold Voltage
Description: This command is used for setting a minimum Threshold Voltage for the
OCP/OPP tests. If the measured voltage is below the threshold voltage
And the OCP/OPP test started the test will not run and an error will be
Flagged.
OPP: START
Command Syntax:
[PRESet] OPP:START {SP}{NR2}{NL}
[PRESet] OPP:START {NL}
Purpose: Set and read the initial value of OPP Test
Description: This command is used for setting the initial value (P-START) of the OPP
Test.
OPP: STEP
Command Syntax:
[PRESet] OPP:STEP {SP}{NR2}{NL}
[PRESet] OPP:STEP {NL}
Purpose: Set and read the increasing value of OPP Test
Description: This command is used for setting the power increments which the
OPP test will follow between the P-START and P-STOP values.
OPP: STOP
Command Syntax:
[PRESet] OPP:STOP {SP}{NR2}{NL}
[PRESet] OPP:STOP {NL}
Purpose: Set and read the maximum value of OPP Test
Description: This command is used to set the maximum power value (P-STOP) of
The OPP test.