50 PRODIGIT
TCONFIG
Command Syntax:
[PRESet:] TONFIG {NORMAL|OCP|OVP|OPP|SHORT}{;NL}
[PRESet:] TONFIG ? {;NL}
Purpose: Set and read the function of Dynamic Test
Description: There are four options of this command. Those are NORMAL mode,
OCP Test, OVP Test, OPP Test and SHORT Mode Test.
STIME
Command Syntax:
[PRESet:] STIME {SP}{NR2}{;NL}
[PRESet:] STIME ? {;NL}
Purpose: Set and read time of the short-circuit test
Description: This command is used for setting the short-circuit test time. If the time
Is set to 0 then there is no time limit. In other words a continuous short
Circuit test will be implemented. If a number is entered then this is the
Test time in milliseconds (ms).
OCP
Command Syntax:
OCP?
Purpose: read OCP testing current.
Description: This command is used for reading the OCP current measured in the OCP
test.
OPP
Command Syntax:
OPP?
Purpose: read OPP testing watt.
Description: This command is used for reading the OPP power for the OPP test.
MPPT
Command Syntax:
[PRESet:] MPPT {SP} ON|OFF{;NL}
Purpose: MPPT (maximum power point tracking) test ON/OFF
Description:
This command is set MPPT test ON / OFF.
MPP
Command Syntax:
[PRESet:] MPP? {;NL}
Purpose:
read MPP
(maximum power point
) data, Read form:
Voltmeter
/
Ammeter/
Power Meter.
Description:
MPP read form: Voltmeter
/ Ammeter/ Power Meter.
BW
Command Syntax:
[PRESet:]BW{SP} {LO│HI} {;│NL}
[PRESet:] BW{?}