PQube User Manual 2.1
Page 15 of 113
For mains connections and current connections, PSL recommends snap-on forked terminals, such as 3M Scotchlok
®
P/N MU18-
6FL or MU14-6FL, or Panduit P10-6lF. For other connections, PSL recommends wire ferrules for stranded wire, such as Panduit
F77 series, for example Panduit F77-6-M.
Figure 1: Your PQube meets all IEC requirements for high-
frequency emissions and susceptibility, both conducted and
radiated. For further protection, you can use clamp-on
ferrites on signal cables to minimize radio-frequency
emissions. For example, these are Panasonic
KRCBC160928B and KRCBC130714B.
Figure 2: To minimize emissions with the optional PS1
Power Supply module, optionally use a shielded power
conductor.
Conductor characteristics
Connection
Minimum wire size
Maximum wire size
Limitations and remarks
Comply with all local safety and installation
requirements and regulations.
PQube terminals
L1, L2, L3, N
[9] [11] [13] [15]
20AWG (0,52 mm
2
)
14 AWG (2.1mm
2
)
Min 600V UL-recognized insulation system required.
These terminals require less than 0,01 amps.
Connection to N (15) is optional. For single phase
monitoring, connect either L1-N or L1-L2 as
appropriate for the mains configuration.
PQube Earth terminal
[32]
Wire size used for
L1,L2,L3,N terminals
14 AWG (2.1mm
2
)
Connect this terminal to a suitable earth connection.
For proper PQube operation, you must connect this
terminal to earth. It is used as a measurement
reference, and as a reference for your PQube’s low
voltage circuits.
Optional
PQube RLY4 terminals
[25] [26]
20AWG (0,52 mm
2
)
14 AWG (2.1mm
2
)
Min 600V UL-recognized insulation system required.
Wire size must be adequate for relay contact load.
These terminals are rated at 30 VAC max, 60 Vdc
max, 2 amps max.
All other PQube
terminals
20AWG (0,52 mm
2
)
14 AWG (2.1mm
2
)
Min 600V UL-recognized insulation system required.
Optional PS1
terminals
20AWG (0,52 mm
2
)
14 AWG (2.1mm
2
)
Min 600V UL-recognized insulation system required.
Shielded cable recommended for minimizing
emissions.
Optional XCT4
terminals
20AWG (0,52 mm
2
)
14 AWG (2.1mm
2
)
Ring lugs or snap-on forked terminals are
recommended for CT connections. Self-retaining
screws in terminals may make ring lugs challenging to
install. Wire size must be adequate for 1-amp or 5-
amp currents. Min 600V UL-recognized insulation
system required
Optional CT4 pass-
through inputs
20AWG (0,52 mm
2
)
Limited by physical
dimension of pass-
through locations:
8.5 mm (0.34”) dia.
Min 600V UL-recognized insulation system required.
Insulated wire is required!