EasyManua.ls Logo

PTW DIADOS E - Problems During Operation

PTW DIADOS E
51 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
DIADOS E
T11035
Operating Manual Error Messages and Troubleshooting
38
D616.131.00/10 en
5.2 Problems during Operation
Problem
Cause
Remedy
No keypad response.
Keypad locked via interface.
Unlock keypad via interface. Refer to interface
description („User Manual RS232 Interface for
PTW-DIADOS E“).
LCD does not illumi-
nate.
The rechargeable batteries
are depleted.
Connect the multi-range power supply. The bat-
teries charge as soon as the power supply is con-
nected.
The calibration data of
the application cannot
be displayed.
The device is not calibrated
for the application.
Download the calibration factors of the appropri-
ate detector via the interface („User Manual
RS232 Interface for PTW-DIADOS E“).

Table of Contents