DIAMENTOR E2
First Time Operation
D651.131.00/05 en 15
4.3 Connecting the Chamber and
Printer and Putting into Operation
Always turn off the device before connecting or
removing peripheral devices.
Connect the chamber to 3 (refer to Figure 3).
Use the original cables from PTW (available in
various lengths up to 40 meters (130 ft)).
Socket SERIAL (Figure 3, item 1): Connect the
printer cable or RS232 interface cable for PC.
Turn on the DIAMENTOR E2 with the On/Off
switch (refer to
Figure 3, item 5). The firmware
version will be displayed for about 2 seconds,
followed by a display test. After completion of
the test the DIAMENTOR E2 is operational.
4.4 Mounting the ionization
chamber
General information
The DIAMENTOR E2 can be operated with every
PTW-DIAMENTOR chamber featuring connecting
system A (
Figure 3, item 3).
The chamber calibration factor must be entered for
the connected chamber (section
4.5). When replac-
ing the chamber, do not forget to enter the new
chamber calibration factor (section
4.5.)
The DIAMENTOR E2 connection cables are avail-
able in various lengths (max. 40 meters (130 ft.)).
Use only the original cables from PTW.