DIAMENTOR E2
Error Messages and Troubleshooting
D651.131.00/05 en 19
6 Error Messages and Troubleshooting
Message Cause Remedy
OFL-Err (Overflow
Error)
The dose area product has exceeded
20 000 000 µGym²
or
the dose area product rate has exceeded
15 000 µGym²/s
Press the RESET button.
Error 51 communication error Turn the device off and on again. When
the message appears again the device
must be repaired.
Error 54 Memory malfunction Turn the device off and on again. When
the message appears again the device
must be repaired.
HU-Err Chamber voltage missing, device is un-
able to measure
Disconnect the chamber cable from the
DIAMENTOR E2. When the message
continues appearing the device is defec-
tive and must be repaired.
When the message clears the chamber
cable is defective and must be replaced.