DIAMENTOR E2
Technical Specifications
D651.131.00/05 en 25
10 Technical Specifications
Only values for which tolerances or limits are specified are guaranteed. Values for which no tolerances are
specified are for information only.
Device designation and use DIAMENTOR E2 T11033
dose area product meter according to IEC 60580 for measurements
in diagnostic radiology
Manufacturer PTW-Freiburg
Display LED, 8 digit seven-segment display; red
Number of measuring channels 1
Measuring principle detection of irradiation by means of ionization chambers, analog
and digital signal processing, digital readout on LED
Operating mode continuous operation
Measuring quantities Dose area product, dose area product rate
Linearity deviation < 2.5 %
Offset current (zero drift) < 0.2 pA at 25°C
Settling time max.2 s
Operational readiness 10 s after power up
Nominal useful range dose area prod-
uct (DAP) rate
(0.01 .... 15 000) µGycm²/s
Dose area product (DAP)
Unit: µGym²
Measuring range: (0.1 ... 20 000 000) µGym²
Accuracy: ± 1 digit
Digital resolution: 0.01 µGym²
Dose area product (DAP) rate
Unit: µGym²/s
Measuring range: (0.1 ... 15 000) µGym²/s
Accuracy: ± 1 digit
Digital resolution: 0.01 µGym²/s