DIAMENTOR E2
For Your Safety
6
D651.131.00/05 en
Safety Information
DANGER
Explosion Hazard – The product is not de-
signed for use in areas of medical locations
where an explosion hazard may occur. An ex-
plosion hazard may result from the use of
flammable anesthetics, skin cleansing agents
and disinfectants. Furthermore, the product is
suitable for use in an oxygen-enriched atmos-
phere only with certain restrictions. The atmos-
phere is oxygen-enriched when the room air
contains more than 25 % oxygen or nitrous
oxide.
WARNING
Shock Hazard – Strictly observe the following
warnings. Failure to do so may endanger the
lives of the patient, the user and other persons
involved.
The product is a system device of
IEC 60601-1 protection class II in
combination with functional earth. It may be
put into operation only when connected to a
properly installed power outlet with earthing
contact.
The installation of the product may only be
performed by trained electro-technical staff.
Before using the product, the user must as-
certain that it is in correct working order and
operating condition.
Inspect the connection cables for damage
before use. Replace damaged cables and
connectors immediately.
When disconnecting the device from the
power line, first remove the connector from
the wall outlet. Then disconnect the cable
from the device.
Devices on which moisture condensation
has developed as a result of temperature
changes must not be switched on unless
completely dry.
Liquids must not enter neither the display
unit nor the adapter. If liquids have entered
the product, it must be thoroughly inspected
before being used again.
Extension cords must not be used.
Operation in the patient environment:
With all peripheral devices connected, the
product represents a medical system. The
product may be operated in the patient envi-
ronment only if
no peripheral devices are connected to the
serial interface or
the peripheral devices connected meet the
requirements IEC 60601-1/
ANSI/AAMI ES 60601-1 /
CAN/CSA-C22.2.No.60601-1
the peripheral devices connected meet the
requirements IEC 60950-1 / UL 60950-1 and
are operated outside the patient environ-
ment.
Exclusion of operation as device with pa-
tient contact:
The device is not for use in direct contact
with the patient. Neither the device nor any
peripheral device may have contact to the
patient.
Never touch the patient and open connec-
tors of the device at the same time.
Exclusion of operation as a controlling in-
strument:
The device is only for use to measure the
dose within radiation field. The product must
not be used to control radiodiagnostic
equipment or radiotherapy units.