User Manual
MULTIDOS
RS232 Interface
D177.131.2/0 2003-07 Di/Se
Question and Answer IconNeed help?

Do you have a question about the PTW MULTIDOS and is the answer not in the manual?

PTW MULTIDOS Specifications

General IconGeneral
BrandPTW
ModelMULTIDOS
CategoryMeasuring Instruments
LanguageEnglish

Summary

1 For your Safety

3 Interface Configuration

5 General Telegrams

5.1 Device Identification

Read device information. x.xx is the firmware version, u describes the radiological unit.

5.5 Error Codes

Lists error codes like E01, E02, E03, E06, E07, E09, E10 and their descriptions.

5.6 Status telegrams

6 Brief Description

Step 2: open communication

Start the communication program on your PC. The program must be set up correctly!

7 Application Dual Channel Dosemeter

7.1 Basic Settings

Set/Read the password protection. Parameter x: 0 = clear protection, 1 = Set protection.

8 Application Multi Channel Dosemeter

8.1 Basic Settings

Set/Read unit of measurement. Parameter x: A = Ampere, S = Gy/s, M = Gy/min, H = Gy/h

9 Application Constancy Check

9.1 Basic Settings

Set/Read unit of measurement. Parameter x: 1 = Ampere, 2 = Gy/s, 3 = Gy/min, 4 = Gy/h

9.5 Errors

10 Application Linear Array LA 48

10.2.2 Single channel data

10.2.3 Data of all channels

11 Application Afterloading