MULTIDOS
RS232 Interface
Application Constancy Check
16 D177.131.2/0
9.3 Correction Telegrams
Telegram Description Response
C;x
Set/Read Air Density Correction 0: Off 1: On C;x
CP;xxxx.x
Set/Read nominal air pressure (500.0 <= p <= 1300.0), default: 1013.2hPa CP;xxxx.x
CT;xx.x
Set/Read the temperature (10.0 <= T <= 40.0), default: 20/22°C CT;xx.x
CD Checking the correction factor temperture/air density without kQE CD;x.xxxx
CK;x.xxxx
Set/Read Beam Quality Factor. (0.0000 < kQE <= 2.0000), default: 1.0000 CK;x.xxxx
9.4 Calibration Factors
Telegram Description Response
XR;s;cc;f Set/Read the radiological calibration factor for channel cc in set s
XR;s;cc;±m.m
mmmE+ee
XRF;s Set/Read the flag for set s for radiologically calibrated/uncalibrated
0: not calibrated, 1: calibrated
XRF;s;x
XRN;s;name Set/Read the name for set s with radiological calibration factors
XRN;s;name
XRD;s;dd.mm
.yyyy
Set/Read the radiological calibration date for set s
XRD;s;dd.mm
.yyyy
XRC;s Renew the check sum radiological calibration factor for set s
XRC;s
9.5 Errors
Telegram Description Response
ERR Checking the error status flag of MULTIDOS
Bit0: MENU_ERROR
Bit1: ELCAL_ERROR
Bit2-7: not used
Bit8: RADCAL_ERROR Set 1
Bit9: RADCAL_ERROR Set 2
Bit10: RADCAL_ERROR Set 3
Bit11: RADCAL_ERROR Set 4
Bit12: RADCAL_ERROR Set 5
Bit13-15: not used
ERR;xxxxx