11
USER MANUAL FOR KRYPTON XG50 THERMAL IMAGING MONOCULAR
• 
Check the eyepiece and the lens and if required remove dust and dirt from the optics (preferably using a 
non-contact method). Cleaning of the exterior surfaces of the optics should only be done with products 
specically designed for this purpose.
• 
Always store the device in its carrying case in a dry, well-ventilated space. For prolonged storage, remove the 
batteries.
TROUBLESHOOTING
The table below lists problems that may occur when using the device. Carry out the recommended checks 
and troubleshooting steps in the order listed in the table. If there are defects not listed in the table or it is 
impossible to resolve the problem yourself, the device should be returned for repair.
Malfunction Possible cause Corrective action
The thermal imager does 
not turn on.
The battery is completely 
discharged
Charge the battery.
The device does not 
operate from an external 
power source.
The USB cable is damaged. Replace the USB cable.
The external power supply is 
discharged.
Charge the external power supply (if necessary).
Blurred image with vertical 
stripes or an uneven 
background.
Calibration is required. Perform image calibration according to the 
Microbolometer Calibration section of the manual.
Poor quality image. There 
is noise or ghost images of 
previous scenes or objects.
Manual calibration has been 
performed with the lens cover 
open.
Check the calibration mode, close the lens cover 
and calibrate the device.
Image is too dark. Brightness or contrast level is 
too low.
Adjust the brightness or contrast.
Color bars appear on 
the display or the image 
disappears.
The device was exposed to static 
charges during operation.
When the exposure to static charges is over, the 
device may either reboot automatically or require 
to be switched off and on again.
The image of the object 
being observed is missing.
You are looking through glass. Remove the glass or change the viewing position 
to avoid it.
Poor image quality / 
reduced detection distance.
These problems may occur during observation in adverse weather conditions 
(snow, rain, fog, etc.).
Smartphone or tablet 
cannot be connected to the 
device.
Device password has been 
changed.
Delete the network and connect again using the 
password saved in the device.
The device is in an area with too 
many Wi-Fi networks that may 
be causing signal interference.
To ensure a stable Wi-Fi connection, relocate the 
device to an area with fewer or no Wi-Fi networks.
Missing or interrupted 
broadcasting via Wi-Fi.
The smartphone or tablet is 
beyond reliable Wi-Fi range. 
There are obstacles between the 
device and the signal receiver 
(e.g. concrete walls).
Move the devices in line-of-sight and within range 
of the Wi-Fi signal.
When the device is used in 
low temperature conditions 
the image quality is 
worse than in positive 
temperatures. 
In positive temperature conditions, objects being observed (surroundings 
and background) heat up differently because of thermal conductivity, thereby 
generating a high temperature contrast. Consequently, the image quality produced 
by the thermal imager will be better. In low-temperature conditions, objects being 
observed (background) will cool down to roughly the same temperature, which 
leads to a greatly reduced temperature contrast and a degraded image quality. This 
is normal for all thermal imaging devices.
Repair of the device is possible within 5 years.