Page 13
P050 Negative End
The point at which, the negative gain line intersects the 0.0m distance y axis.
This parameter determines the angle at which the rate of near signal amplitude
suppression, (gain attenuation), is applied, over a distance starting at the gain
breakpoint . See the explanation of P43 for details.
Values Min = -999
Max = 999
P051 Margin Enable
This parameter is associated with P52 End Margin.
1 switches the feature on
0 (off) is the default value
Values Min = 0
Max = 1
Default = 80
Default = 0 off