What to do if my Pulsar ULTRA TWIN display an incorrect reading for the current level?
- KKevin GuerreroAug 17, 2025
If your Pulsar Measuring Instruments are displaying an incorrect reading, measure the actual distance from the transducer head to the surface of the material. Then, enter Program Mode and directly access P21 (Set Distance), type in the measured distance, press ENTER, and ENTER again when prompted. Wait until SET is displayed and return to Run Mode. The display should now update to the correct reading.