Page 162 
Failsafe 
P1 or P2 
P1-808, P2-808 Fail-safe Mode 
By default, if a fail-safe condition occurs, then the display, relays and the 
mA  output  are  held  at  their  last  known  values  until  a  valid  reading  is 
obtained. 
If required, then you can change this so that the unit goes to high (100% of 
span), or low (empty) as follows: 
Remain at the last known value 
Will fail to the high value (100% of Span). 
Will fail to the low value (empty) 
 
See Also P*218, P*228, P*238, P*248, P*258, P*268 - Relay Fail-safe and P*840 – mA 1 Output 
Fail-safe and P*898 – mA 2 Fail-safe. 
Important Information 
In the event of a fail-safe condition occurring, the displays, relays and mA 
Outputs can be configured to fail to a condition which is independent of 
each other. To set independent Relay Failsafe see P*218, P*228, P*238, 
P*248, P*258, P*268. For independent mA 1 Output Failsafe see P*840 
and mA 2 Output Failsafe see P*898. 
 
 
P1-809, P2-809 Fail-safe Time 
In the event of a fail-safe condition the fail safe timer determines the time 
before fail-safe mode is activated. Default = 2mins 
If the timer activates, the unit goes into fail-safe, as determined by P1-808, 
P2-808 (Display), P*218, 228, 238, 248, 258, 268 (Relays), P*840 (mA 1 
Output) and P*898 (mA 2 Output). When this happens, you will see the 
message  “Failed  Safe!”  on  the  display,  along  with  a  message  explaining 
why (lost echo or transducer fault, for example) 
When a valid measurement is obtained then the display(s), relays and mA 
output(s) will be restored and the timer is reset.