OPERATION
Flushing flexible endoscopes with the FlexiPump
Independent Flushing System
10
If the pump sounds or looks stressed,
turn the speed setting down to avoid
unnecessary wear on the motor.
PSI Settings
Please refer to the manufacturer’s IFU for each endoscope’s pressure requirements. The FlexiPump system
can be used with a low-pressure tube set or a higher-pressure tube set. In general, smaller flexible devices
(urology scopes and bronchoscopes, for example) require lower pressure.
General rules and guidelines for flexible endoscope speed control
There are three (3) tubes on each FlexiPump tube set. If any tubes are not in use, be sure they are
secured with end caps provided in the tube set.
When using all three tubes of the tube set for flushing, the MEDIUM or HIGH FlexiPump speed
should be used.
When using two tubes, make sure the third tube is capped and the FlexiPump is set for LOW,
MEDIUM or (in some cases) HIGH speed.
When using only one of the three tubes of the FlexiPump tube set, cap off the other two tubes and
setthe FlexiPump speed to LOW, MEDIUM or (in rare cases) HIGH speed.
PLEASE NOTE: the FlexiPump Independent Flushing System is intended for pre-cleaning and flushing of lumened
devices. It is not for final treatment and does not replace manual brushing and rinsing. Refer to each medical device
manufacturer’s instructions for use for complete pre-cleaning instructions.
FlexiPump™ Independent Flushing System User Manual