MAINTENANCE
Decontaminating the tube set
12
Bleach or alcohol
disinfectants are
not compatible
with the FlexiPump
Disinfect the tube sets at least daily, preferably at every shift change.
Step 1 Pour 500mL of your ready to use disinfectant into an open, clean container.
Step 2 Place the drop down tube into the container with disinfectant, assuring it's fully submerged.
Step 3
Press the forward flush button to begin filling tube set with disinfectant. Run at least 100mL of disinfectant
through the tube leads. Press STOP.
Step 4
Attach end caps onto all the tube leads and set the speed control on Low. Position your pressure relief valves
over a container or sink bay. Run a Forward Flush until disinfect is expelled through the pressure relief valves. Press
STOP. Your tubes are now completely filled with disinfectant.
Step 5
The disinfectant should remain inside tubing for the contact time of the disinfectant.
Step 6 After the contact time has elapsed, in a empty sink basin, point the tube leads downward. Slowly take off the
end caps. Some pressure will have built up inside, so use caution.
Step 7 Run the remainder of the disinfectant through the tube sets.
Step 8 Fill a container with 500 mL of clean water and place the drop down tube inside. Rinse at least 500mL
through the tube leads to ensure a proper rinse.
Cleaning FlexiPump surfaces
The surfaces of the system should be cleaned daily and whenever there is visible soil.
Step 1 Wipe all external surfaces with a ready to use disinfectant, including the tubing, and allow
them to air-dry.
Step 2 Wipe the surfaces of the Nest accessory and/or Rail accessory (if used) and let them air-dry.