Appendix D Calibration (Using Generator Service Tools)
Revision W HF Series X-ray Generators - Service Manual
D-64 Quantum Medical Imaging, LLC
5. The Calibration wizard - Copy Receptor Calibration screen (Step 3.) is dis-
played (see below). To perform another copy operation, press Yes but-
ton and repeat steps above. To exit the copy utility, press No button.
Figure D-35. Calibration Wizard - Copy Receptor Calibration (Step 3)