Page #33 background imageLoading...
Page #33 background image
- 33 -
9.1. USERS
Each user is characterized by the following data:
NAME (30 characters), CODE (6 characters),
PASSWORD (8 characters, digits only),
ACCESS (USER, ADVANCED, ADMIN),
LANGUAGE (any of the available).
Access levels
The balance software has three access levels: USER, ADVANCED, ADMINISTRATOR.
Once the balance is switched on, the display stays active all the time, this enables carrying out mass
measurements even when no user is logged-in.
User parameters, databases and software functions may be edited depending on the access
level granted to a particular user. Access levels are presented in the table below.
Access
levels
Permissions
USER
Access to parameters of the submenu <Reading>, permission to change settings for
param
eter group <Other> except for <Date and Time>. Permission to start and carry
out all the weighing processes. You
can preview information in <Databases> and
define universal variables.
ADV
Access to editing parameters of the submenu: <Reading>; <Working modes>;
<Communication>; <Devices>; <Other> except for <Date and Time>. Permission to
start and carry out all the weighing processes.
ADMIN
Access to all user parameters and functions, permission to edit databases.
In order to add a user, follow the scheme shown below, add a user and assign a name to him/her.
Once the user name is added, enter the following:
User code Max 6 characters,
User password Max 8 numbers.
Next select:
Access level (USER, ADV, ADMIN)
Language
In order to remove a user you should follow this procedure:
Enter the user database
Select the user that is to be removed from the list
Press button

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the RADWAG AS R2 and is the answer not in the manual?

RADWAG AS R2 Specifications

General IconGeneral
Readability0.1 mg
DisplayLCD (backlit)
Tare Range-220 g
Linearity±0.2 mg
Protection ClassIP43
CalibrationInternal
InterfaceRS232, USB-A, USB-B
Power Supply100 ÷ 240 V AC 50 / 60 Hz
Maximum Capacityup to 220 g