4.
Go to Device Manager and check number assigned to
virtual COM port.
For this very case it is
STMicroelectronics Virtual
COM Port (COM8).
5.
Set balance parameters: select USB value for parameter COMPUTER/PORT.
6.
Run software for measurement readout.
7.
Use the program to set parameters for communication select respective COM port (for the
very case it is COM8), i.e. the one which has been formed while installing drivers.
8.
Start cooperation.
13. PERIPHERAL DEVICES
PERIPHERAL DEVICES menu is comprised within Parameters menu. It is accessed by pressing
key. The menu features list of devices that can cooperate with the balance.
13.1. COMPUTER
<COMPUTER> submenu allows the user to select a port for connection of a computer running
software which enables:
communication with a balance,
switching the continuous transmission on and off, and
cooperation of a balance with E2R software of Radwag production.
13.1.1. Computer Connection Port
Procedure:
- press button

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the RADWAG AS R2 and is the answer not in the manual?

RADWAG AS R2 Specifications

General IconGeneral
Readability0.1 mg
DisplayLCD (backlit)
Tare Range-220 g
Linearity±0.2 mg
Protection ClassIP43
CalibrationInternal
InterfaceRS232, USB-A, USB-B
Power Supply100 ÷ 240 V AC 50 / 60 Hz
Maximum Capacityup to 220 g