- 66 -
It is possible to print out the report again by pressing
button.
To exit this window, press
button. The main window for <TOTALISING> mode is displayed,
data related to carried out measurements is automatically zeroed.
11.12. PEAK HOLD
This function enables snapping peak hold applied to the weighing pan during a single checkweighing
process. Apart from standard settings for this mode (described in the weighing mode), additional
setting for threshold values activating function has been introduced.
The option is available in the settings for mode <PEAK HOLD> and it is:
THRESHOLD this function specifies starting point for control of maximum pressure applied
onto the weighing pan, wherein the said control is performed by the balance software. The
threshold must be set in accordance with the needs prior to the measuring process.
F Shortcut Keys
You can declare which particular weighing function is to be launched using F1, F2, F3 or F4 keys.
For procedure informing you how to declare the particular function, read F Shortcut Keys section.
11.12.1. Means of Operation
Enter <PEAK HOLD> mode
Once you select the mode, the function is active and information about the net mass is
displayed in a bottom line (only if a different information has not been selected by a user).
In order to provide proper operation, set the threshold in grams determining point beyond which
the function starts to register maximum force apllied.
From now on the balance registers and holds every single weighing which is above the
threshold, and which is higher than the result of the previous peak hold. If the software detects
mass above the threshold, the highest detected indication is held on the main display and the
pictogram <Max> is shown at the top of the display.
You can print the result out by pressing
button.
The start of the next process of peak hold measurement is possible only after removing the
load from the weighing pan and pressing
button. This causes returning to the main
window of <PEAK HOLD> mode, pictogram <Max> is automatically deleted.
CAUTION!
Current unit may be selected only if the weighing result is not snapped. In order to select the unit, Units/Esc
button is used. If any result has already been snapped than Esc button deletes the last snapped Max result.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the RADWAG AS R2 and is the answer not in the manual?

RADWAG AS R2 Specifications

General IconGeneral
Readability0.1 mg
DisplayLCD (backlit)
Tare Range-220 g
Linearity±0.2 mg
Protection ClassIP43
CalibrationInternal
InterfaceRS232, USB-A, USB-B
Power Supply100 ÷ 240 V AC 50 / 60 Hz
Maximum Capacityup to 220 g