- 86 -
Immediately send measurement result in a current measuring unit
Format: SUI CR LF
Response options:
SUI_I CR LF
- command understood but not accessible at this moment
MASS FRAME
- immediate response: mass value in a current measuring unit
Response format:
1
2
3
4
5
6
7-15
16
17
18
19
20
21
S U I
stability
marker
space character mass space unit CR LF
An example:
S U I CR LF
command sent from a computer
S U I ? _ - _ _ _ 5 8 . 2 3 7 _ k g _ CR LF
- command carried out, immediate response: mass value
in a current measuring unit
where: _ - space
Switch on continuous transmission in a basic measuring unit
Format: C1 CR LF
Response options:
C1_I CR LF
- command understood but not accessible at this moment
C1_A CR LF
- command understood and in progress
MASS FRAME
- response: mass value in a basic measuring unit
Response format:
1
2
3
4
5
6
7-15
16
17
18
19
20
21
S I space
stability
marker
space character mass space unit CR LF
Switch off continuous transmission in a basic measuring unit
Format: C0 CR LF
Response options:
C0_I CR LF
- command understood but not accessible at this moment
C0_A CR LF
- command understood and carried out
Switch on continuous transmission in a current measuring unit
Format: CU1 CR LF
Response options:
CU1_I CR LF
- command understood but not accessible at this moment
CU1_A CR LF
- command understood and in progress
MASS FRAME
- response: mass value in a current measuring unit
Response format:
1
2
3
4
5
6
7-15
16
17
18
19
20
21
S U I
stability
marker
space character mass space unit CR LF
Switch off continuous transmission in a current measuring unit
Format: CU0 CR LF
Response options:
CU0_I CR LF
- command understood but not accessible at this moment
CU0_A CR LF
- command understood and carried out
Set min checkweighing threshold
Format: DH_XXXXX CR LF, where: _ - space, XXXXXmass format
Response options:
DH_OK CR LF
- command carried out
ES CR LF
- command not recognised (mass format incorrect)

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the RADWAG AS R2 and is the answer not in the manual?

RADWAG AS R2 Specifications

General IconGeneral
Readability0.1 mg
DisplayLCD (backlit)
Tare Range-220 g
Linearity±0.2 mg
Protection ClassIP43
CalibrationInternal
InterfaceRS232, USB-A, USB-B
Power Supply100 ÷ 240 V AC 50 / 60 Hz
Maximum Capacityup to 220 g