- 138 -
Start the proces.
Following the displayed messages set the correct options.
Enter the ambient temperature, next press ENTER for confirmation.
The computer software proceeds to the next step, i.e. ambient
humidity settings.
Set the ambient humidity, next press ENTER for confirmation. The
computer software proceeds to the next step, i.e. air pressure
settings.
Set the air pressure, next press ENTER for confirmation. The
computer software proceeds to the next step, i.e. entering the first
control volume (V1) for a tested pipette.