- 139 -
Enter (V1) control volume. For fixed volume pipettes (P2.13.5
VOLUME DETERMINATION parameter set to value <1>) this is the
one and only value to be entered. The computer software proceeds to
the next step, i.e. entering the second control volume (V2) for a tested
pipette.
Enter (V2) control volume. The computer software proceeds to the
next step, i.e. entering the third control volume (V3) for a tested
pipette.
Enter (V3) control volume.
Upon entering the data the computer software proceeds to pipettes
calibration.
Follow the displayed description in order to complete the procedure.
By means of pipette dose the first portion of water, wait until the
measurement is stable and press ENTER to confirm.